Development and Validation of an LC-MS/MS Method for Simultaneous Determination of Short Peptide-Based Drugs in Human Blood Plasma

Molecules. 2022 Nov 14;27(22):7831. doi: 10.3390/molecules27227831.

Abstract

A novel HPLC-ESI-MS/MS method for simultaneous gonadotropin-releasing hormone (GnRH) analogs and somatostatin analog quantitation was developed and validated. The developed method was successfully applied to pharmacokinetic studies. The sample preparation process included solid-phase extraction (SPE). Effective chromatographic separation of the analytes and internal standard (dalargin) was achieved with a C18 column, using a gradient elution with two mobile phases: 0.1% v/v formic acid (aqueous solution) and 0.1% v/v formic acid (acetonitrile solution). The linearity of the method was demonstrated within a concentration range of 0.5-20 ng/mL, with correlation coefficients between 0.998-0.999 for goserelin, buserelin, triptorelin, and octreotide, respectively. The relative standard deviation (RSD, %) values for method accuracy and precision did not exceed 20% at the lower level of quantitation (LLOQ) or 15% at other concentration levels.

Keywords: GnRH analogs; LC-ESI-MS/MS; peptide drugs; solid-phase extraction; somatostatin analogue.

MeSH terms

  • Chromatography, Liquid / methods
  • Humans
  • Peptides
  • Plasma*
  • Reproducibility of Results
  • Tandem Mass Spectrometry* / methods

Substances

  • formic acid
  • Peptides

Grants and funding

This research received no external funding.