Goal: This work presents the development of a Hybrid Mock Circulatory Loop (HMCL) to simulate hemodynamics at patient-specific level in terms of both 3D geometry and inlet/outlet boundary conditions.

Methods: Clinical data have been processed to define the morphological and functional patient-specific settings. A piston pump is used to impose a parametric flow rate profile at the inlet of the hemodynamic circuit. In order to guarantee the physiological pressure and flow conditions, a specific hybrid chamber system including a real-time control has been designed and implemented. The developed system was validated firstly in a single outlet branch model and, secondly, on a 3D printed patient-specific multi-branch phantom. Finally, for the 3D phantom, the outlet flow profiles were compared with the corresponding in-vivo flow data.

Results: Results showed that the root mean squared error between the prescribed setpoint and the measured pressures was always below 3 mmHg (about 2.5%) for all cases. The obtained flow profiles for the patient-specific model were in agreement with the related functional in-vivo data.

Significance: The capability to reproduce physiological hemodynamics condition, with high-fidelity, plays a significant role in the cardiovascular research. The developed platform can be used to assess the performances of cardiovascular devices, to validate numerical simulations, and to test imaging systems.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Hemodynamics / physiology
  • Humans
  • Hydrodynamics*
  • Models, Cardiovascular*