Biomechanical Properties of Novel Lateral Hole Pedicle Screws and Solid Pedicle Screws: A Comparative Study in the Beagle Dogs

Orthop Surg. 2023 Jan;15(1):328-336. doi: 10.1111/os.13596. Epub 2022 Nov 21.

Abstract

Objective: Although pedicle screws are widely used to reconstruct the stability of the spine, screw loosening is a common complication after spine surgery. The main objective of this study was to investigate whether the application of the hollow lateral hole structure had the potential to improve the stability of the pedicle screw by comparing the biomechanical properties of the novel lateral hole pedicle screws (LHPSs) with those of the solid pedicle screws (SPSs) in beagle dogs.

Methods: The cancellous bone of the distal femur, proximal femur, distal tibia, and proximal tibia were chosen as implantation sites in beagle dogs. In each of 12 dogs, four LHPSs, and four SPSs were implanted into both lower limbs. At 1, 2, and 3 months after surgery, four dogs were randomly sampled and sacrificed. The LHPS group and SPS group were subdivided into four subgroups according to the length of their duration of implantation (0, 1, 2, 3 months). The biomechanical properties of both pedicle screws were evaluated by pull-out and the cyclic bending tests.

Results: The results of the study showed that no significant difference was found between LHPSs (276.62 ± 50.11 N) and SPSs (282.47 ± 42.98 N) in pull-out tests at time 0 (P > 0.05). At the same time point after implantations, LHPSs exhibited significantly higher maximal pullout strength than SPSs (month 1: 360.51 ± 25.63 vs 325.87 ± 28.11 N; month 2: 416.59 ± 23.78 vs 362.12 ± 29.27 N; month 3: 447.05 ± 38.26 vs 376.63 ± 32.36 N) (P < 0.05). Moreover, compared with SPSs, LHPSs withstood more loading cycles (month 2: 592 ± 21 vs 534 ± 48 times; month 3: 596 ± 10 vs 543 ± 59 times), and exhibiting less displacement before loosening at month 2 (1.70 ± 0.17 vs 1.96 ± 0.10 mm) and 3 (1.69 ± 0.19 vs 1.92 ± 0.14 mm) (P < 0.05), but no significant difference in time 0 and month 1 (P > 0.05).

Conclusions: The pedicle screw with the hollow lateral hole structure could allow bone to grow into the inner architecture, which improved biomechanical properties by extending the contact area between screw and bone tissue after implantation into the cancellous bone. It indicated that LHPS could reduce loosening of the pedicle screws in long term after surgery.

Keywords: Animal model; Cyclic bending test; Pedicle screw; Pull-out test; Screw loosening.