Trabecular bone microstructure analysis on data from a novel twin robotic X-ray device

Acta Radiol. 2023 Apr;64(4):1566-1572. doi: 10.1177/02841851221134973. Epub 2022 Nov 13.

Abstract

Background: Bone strength is related to both mineral density (BMD) and the bone microstructure. However, only the assessment of BMD is available in clinical routine care today.

Purpose: To analyze and study the correlation of trabecular bone microstructure from the imaging data of a prototype Multitom Rax system, using micro-computed tomography (CT) data as the reference method (Skyscan 1176).

Material and methods: Imaging data from 14 bone samples from the human radius were analyzed regarding six bone structure parameters, i.e. trabecular nodes, separation, spacing, and thickness as well as bone volume (BV/TV) and structural model index (SMI).

Results: All six structure parameters showed strong correlations to micro-CT with Spearman correlation coefficients in the range of 0.83-0.93. BV/TV and SMI had a correlation >0.90. Two of the parameters, namely, separation and number, had mean values in the same range as micro-CT. The other four were either over- or underestimated.

Conclusion: The strong correlation between micro-CT and the clinical imaging system, indicates the possibility for analyzing bone microstructure with potential to add value in fracture assessment using the studied device in a clinical workflow.

Keywords: Trabecular bone; bone microstructure; imaging; osteoporosis; segmentation.