Shielding device for endoscopic procedures during lower gastrointestinal endoscopy

DEN Open. 2022 Oct 10;3(1):e173. doi: 10.1002/deo2.173. eCollection 2023 Apr.

Abstract

Objectives: The coronavirus pandemic significantly impacted endoscopic practice. During lower gastrointestinal endoscopy, infectious substances disseminate; therefore, we developed an infection control device (STEP-L) for lower gastrointestinal endoscopy and examined its usefulness.

Methods: STEP-L wraps around the patient's buttocks and covers the endoscope. Using lower endoscopy training models, three endoscopists performed 18 colonoscopies with STEP-L (group S) and without (group C). Endoscopic insertion time and pigmented areas of ​​gloves and diapers after the examination were compared between both groups.

Results: Insertion of the endoscope up to the cecum was possible in all 18 examinations. The insertion time to the cecum was 52.4 ± 19.0 s in group S and 53.9 ± 13.3 s in group C. The pigmented areas of the ​​gloves measured 39,108.0 ± 16,155.3 pixels in group C, but were significantly reduced to 2610.5 ± 4333.8 pixels in group S (p < 0.05). The pigmented areas of the diapers measured 2280.9 ± 3285.2 pixels in group C, but were significantly reduced to 138.0 ± 82.9 pixels in group S (p < 0.05).

Conclusions: Using STEP-L does not change the insertion time, and is technically feasible. STEP-L significantly reduces the adhesion of virtual pollutants to the surroundings, suggesting that this device is useful for infection control during lower gastrointestinal endoscopy.

Keywords: aerosol infection; colonoscopy; contact infection; coronavirus; droplet infection.