Electrolyte Analysis in Blood Serum by Laser-Induced Breakdown Spectroscopy Using a Portable Laser

Molecules. 2022 Sep 29;27(19):6438. doi: 10.3390/molecules27196438.

Abstract

The fast and reliable analysis of electrolytes such as K, Na, Ca in human blood serum has become an indispensable tool for diagnosing and preventing diseases. Laser-induced breakdown spectroscopy (LIBS) has been demonstrated as a powerful analytical technique on elements. To apply LIBS to the quantitative analysis of electrolyte elements in real time, a self-developed portable laser was used to measure blood serum samples supported by glass slides and filter paper in this work. The partial least squares regression (PLSR) method was employed for predicting the concentrations of K, Na, Ca from serum LIBS spectra. Great prediction accuracies with excellent linearity were obtained for the serum samples, both on glass slides and filter paper. For blood serum on glass slides, the prediction accuracies for K, Na, Ca were 1.45%, 0.61% and 3.80%. Moreover, for blood serum on filter paper, the corresponding prediction accuracies were 7.47%, 1.56% and 0.52%. The results show that LIBS using a portable laser with the assistance of PLSR can be used for accurate quantitative analysis of elements in blood serum in real time. This work reveals that the handheld LIBS instruments will be an excellent tool for real-time clinical practice.

Keywords: blood serum; electrolytes; laser-induced breakdown spectroscopy; partial least squares regression.

MeSH terms

  • Electrolytes
  • Humans
  • Lasers*
  • Least-Squares Analysis
  • Serum*
  • Spectrum Analysis / methods

Substances

  • Electrolytes