Voluntary Field Recall of Advanced Bionics HiRes Cochlear Implants: A Single-Institution Experience

Otol Neurotol. 2022 Dec 1;43(10):e1094-e1099. doi: 10.1097/MAO.0000000000003711. Epub 2022 Oct 6.

Abstract

Objectives: In 2020, Advanced Bionics (AB) announced a recall of two cochlear implant (CI) models, the "HiRes Ultra" and "HiRes Ultra 3D", because of reports of hearing degradation. The present study examines clinical parameters and patient features in cases of device failure and evaluates outcomes after reimplantation.

Materials and methods: A series of 52 patients implanted with the recalled devices experienced suspected device failure and subsequently underwent revision CI placement at a tertiary academic medical center between December 2019 and November 2021.

Results: Consonant-nucleus-consonant scores and individual phonemes increased significantly between patients' preoperative evaluation and primary cochlear implantation. Performance declined significantly before revision and recovered after revision CI placement. Similarly, pure-tone average thresholds improved between preoperative and primary CI, fell before revision surgery, and were corrected with revision implantation. As a group, patients reached their peak hearing performance significantly faster after revision CI (mean ± standard deviation, 53.4 ± 51.8 d) compared with their primary CI (mean ± standard deviation, 260.6 ± 245.9 d). Electrical field imaging performed by AB and device impedance measurements were found to be abnormal in the basally positioned electrodes (electrodes 9-16).

Conclusion: Hearing performance degradation is significant in AB Ultra device failures and seems to be linked to the basal-most electrodes in the array. Revision outcomes have been robust, necessitating continued monitoring of affected patients and support for reimplantation procedures.

Level of evidence: IV.

MeSH terms

  • Bionics
  • Cochlear Implantation*
  • Cochlear Implants*
  • Hearing Tests
  • Humans
  • Reoperation
  • Retrospective Studies
  • Speech Perception*