A C-arm photon counting CT prototype with volumetric coverage using multi-sweep step-and-shoot acquisitions

Phys Med Biol. 2022 Oct 17;67(21):10.1088/1361-6560/ac950d. doi: 10.1088/1361-6560/ac950d.

Abstract

Objective.Existing clinical C-arm interventional systems use scintillator-based energy-integrating flat panel detectors (FPDs) to generate cone-beam CT (CBCT) images. Despite its volumetric coverage, FPD-CBCT does not provide sufficient low-contrast detectability desired for certain interventional procedures. The purpose of this work was to develop a C-arm photon counting detector (PCD) CT system with a step-and-shoot data acquisition method to further improve the tomographic imaging performance of interventional systems.Approach.As a proof-of-concept, a cadmium telluride-based 51 cm × 0.6 cm PCD was mounted in front of a FPD in an Artis Zee biplane system. A total of 10 C-arm sweeps (5 forward and 5 backward) were prescribed. A motorized patient table prototype was synchronized with the C-arm system such that it translates the object by a designated distance during the sub-second rest time in between gantry sweeps. To evaluate whether this multi-sweep step-and-shoot acquisition strategy can generate high-quality and volumetric PCD-CT images without geometric distortion artifacts, experiments were performed using physical phantoms, a human cadaver head, and anin vivoswine subject. Comparison with FPD-CT was made under matched narrow beam collimation and radiation dose conditions.Main results.Compared with FPD-CT images, PCD-CT images had lower noise and improved visualization of low-contrast lesion models, as well as improved visibility of small iodinated blood vessels. Fine structures were visualized more clearly by the PCD-CT than the highest-available resolution provided by FPD-CBCT and MDCT. No perceivable geometric distortion artifacts were observed in the multi-planar PCD-CT images.Significance.This work is the first demonstration of the feasibility of high-quality and multi-planar (volumetric) PCD-CT imaging with a rotating C-arm gantry.

Keywords: CT image quality; CT systems; Photon counting CT.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Cone-Beam Computed Tomography* / methods
  • Humans
  • Phantoms, Imaging
  • Tomography, X-Ray Computed* / methods