Background: The Carina system (Cochlear Ltd., Sydney, Australia) is a totally implantable device providing acoustic amplification in adult patients with moderate-to-severe sensorineural or mixed hearing loss. One of the main concerns about such a totally implantable device has been represented by the subcutaneous battery lifespan. The aim of this article is to report the analysis of battery performances in a series of Carina-implanted patients after a long follow up.
Methods: In this retrospective study, the technical data of a series of patients implanted with the Carina middle ear implant in our clinic have been analysed, extracting the data from the log of telemetric measures.
Results: The mean lifespan cutback was 0.43 h/years (from 0 to 0.71 h/year), with a strong negative significant correlation between the follow-up period and the percentage of battery residual lifespan.
Conclusion: The lifespan of the Carina's battery seems consistent with the manufacturer statement of a pluri-decennial lifespan, avoiding the need of an early surgical substitution and providing a full day of use of the system even after up to 12 years from the implantation.
Keywords: active middle ear implant; battery; totally implantable hearing aid.