Remote self-collection of capillary blood using upper arm devices for autoantibody analysis in patients with immune-mediated inflammatory rheumatic diseases

RMD Open. 2022 Sep;8(2):e002641. doi: 10.1136/rmdopen-2022-002641.

Abstract

Objectives: To evaluate the feasibility, accuracy, usability and acceptability of two upper arm self-sampling devices for measurement of autoantibodies and C reactive protein (CRP) levels in patients with immune-mediated rheumatic diseases (IMRDs).

Methods: 70 consecutive patients with IMRD with previously documented autoantibodies were assigned to supervised and unsupervised self-collection of capillary blood with the Tasso+ or TAP II device. Interchangeability of 17 biomarkers with standard venesection was assessed by: concordance, correlation, paired sample hypothesis testing and Bland-Altman plots. Patients completed an evaluation questionnaire, including the System Usability Scale (SUS) and Net Promoter Score (NPS).

Results: While 80.0% and 77.0% were able to safely and successfully collect capillary blood using the Tasso+ and TAP II within the first attempt, 69 of 70 (98.6%) patients were successful in collecting capillary blood within two attempts. Concordance between venous and capillary samples was high; 94.7% and 99.5% for positive and negative samples, respectively. For connective tissue disease screen, anti-Ro52 and anti-proteinase 3 autoantibody levels, no significant differences were observed. Self-sampling was less painful than standard venesection for the majority of patients (Tasso+: 71%; TAP II: 63%). Both devices were well accepted (NPS; both: +28%), usability was perceived as excellent (SUS; Tasso+: 88.6 of 100; TAP II: 86.0 of 100) and 48.6 %/62.9% of patients would prefer to use the Tasso+/TAP II, respectively, instead of a traditional venous blood collection.

Conclusions: Remote self-collection of capillary blood using upper arm-based devices for autoantibody and CRP analysis in patients with autoimmune rheumatic diseases is feasible, accurate and well accepted among patients.

Trial registration number: WHO International Clinical Trials Registry (DRKS00024925).

Keywords: Antibodies, Anticardiolipin; Autoimmune Diseases; Health services research.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Arm
  • Autoantibodies
  • Autoimmune Diseases*
  • Humans
  • Rheumatic Diseases* / diagnosis
  • Surveys and Questionnaires

Substances

  • Autoantibodies

Associated data

  • DRKS/DRKS00024925