Five-Year Outcomes in Patients With Fully Magnetically Levitated vs Axial-Flow Left Ventricular Assist Devices in the MOMENTUM 3 Randomized Trial

JAMA. 2022 Sep 27;328(12):1233-1242. doi: 10.1001/jama.2022.16197.

Abstract

Importance: Although durable left ventricular assist device (LVAD) therapy has emerged as an important treatment option for patients with advanced heart failure refractory to pharmacological support, outcomes, including survival, beyond 2 years remain poorly characterized.

Objective: To report the composite end point of survival to transplant, recovery, or LVAD support free of debilitating stroke (Modified Rankin Scale score >3) or reoperation to replace the pump 5 years after the implant in participants who received the fully magnetically levitated centrifugal-flow HeartMate 3 or axial-flow HeartMate II LVAD in the MOMENTUM 3 randomized trial and were still receiving LVAD therapy at the 2-year follow-up.

Design, setting, and participants: This observational study was a 5-year follow-up of the MOMENTUM 3 trial, conducted in 69 US centers, that demonstrated superiority of the centrifugal-flow LVAD to the axial-flow pump with respect to survival to transplant, recovery, or LVAD support free of debilitating stroke or reoperation to replace the pump at 2 years. A total of 295 patients were enrolled between June 2019 to April 2021 in the extended-phase study, with 5-year follow-up completed in September 2021.

Exposures: Of 1020 patients in the investigational device exemption per-protocol population, 536 were still receiving LVAD support at 2 years, of whom 289 received the centrifugal-flow pump and 247 received the axial-flow pump.

Main outcomes and measures: There were 10 end points evaluated at 5 years in the per-protocol population, including a composite of survival to transplant, recovery, or LVAD support free of debilitating stroke or reoperation to replace the pump between the centrifugal-flow and axial-flow pump groups and overall survival between the 2 groups.

Results: A total of 477 patients (295 enrolled and 182 provided limited data) of 536 patients still receiving LVAD support at 2 years contributed to the extended-phase analysis (median age, 62 y; 86 [18%] women). The 5-year Kaplan-Meier estimate of survival to transplant, recovery, or LVAD support free of debilitating stroke or reoperation to replace the pump in the centrifugal-flow vs axial-flow group was 54.0% vs 29.7% (hazard ratio, 0.55 [95% CI, 0.45-0.67]; P < .001). Overall Kaplan-Meier survival was 58.4% in the centrifugal-flow group vs 43.7% in the axial-flow group (hazard ratio, 0.72 [95% CI, 0.58-0.89]; P = .003). Serious adverse events of stroke, bleeding, and pump thrombosis were less frequent in the centrifugal-flow pump group.

Conclusions and relevance: In this observational follow-up study of patients from the MOMENTUM 3 randomized trial, per-protocol analyses found that receipt of a fully magnetically levitated centrifugal-flow LVAD vs axial-flow LVAD was associated with a better composite outcome and higher likelihood of overall survival at 5 years. These findings support the use of the fully magnetically levitated LVAD.

Trial registration: ClinicalTrials.gov Identifier: NCT02224755 and NCT03982979.

Publication types

  • Comparative Study
  • Multicenter Study
  • Observational Study
  • Randomized Controlled Trial

MeSH terms

  • Female
  • Follow-Up Studies
  • Heart Failure / surgery*
  • Heart-Assist Devices* / adverse effects
  • Humans
  • Kaplan-Meier Estimate
  • Male
  • Middle Aged
  • Stroke / etiology
  • Treatment Outcome

Associated data

  • ClinicalTrials.gov/NCT03982979
  • ClinicalTrials.gov/NCT02224755