Beamline simulations using monochromators with high d-spacing crystals

J Synchrotron Radiat. 2022 Sep 1;29(Pt 5):1157-1166. doi: 10.1107/S160057752200707X. Epub 2022 Aug 12.

Abstract

Monochromators for synchrotron radiation beamlines typically use perfect crystals for the hard X-ray regime and gratings for soft X-rays. There is an intermediate range, typically 1-3 keV (tender X-rays), which common perfect crystals have difficulties covering and gratings have low efficiency, although some less common crystals with high d-spacing could be suitable. To evaluate the suitability of these crystals for a particular beamline, it is useful to evaluate the crystals' performance using tools such as ray-tracing. However, simulations for double-crystal monochromators are only available for the most used crystals such as Si, Ge or diamond. Here, an upgrade of the SHADOW ray-tracing code and complementary tools in the OASYS suite are presented to simulate high d-spacing crystals with arbitrary, and sometimes complex, structures such as beryl, YB66, muscovite, etc. Isotropic and anisotropic temperature factors are also considered. The YB66 crystal with 1936 atomic sites in the unit cell is simulated, and its applicability for tender X-ray monochromators is discussed in the context of new low-emittance storage rings.

Keywords: OASYS; SHADOW; YB66; crystal monochromators; high d-spacing crystal; ray tracing.