All-optical thermometry using a single multimode fiber endoscope and diamond nanoparticles containing nitrogen vacancy centers

Rev Sci Instrum. 2022 Aug 1;93(8):083705. doi: 10.1063/5.0102531.

Abstract

Photoluminescence (PL) spectra from diamond nanoparticles containing negative nitrogen vacancy centers were measured by using a single multimode fiber endoscope combined with a high-sensitivity spectroscopy system. A laser light spot was produced at the distal end of the endoscope and the PL spectra from a temperature-controlled ensemble of diamond nanoparticles were measured. After calibrating the sensitivity and wavelength of the spectroscopy system, the temperature dependence of the zero-phonon line peak wavelength similar to those previously reported was obtained.