Reducing cardiac implantable electronic device-induced artefacts in cardiac magnetic resonance imaging

Eur Radiol. 2023 Feb;33(2):1229-1242. doi: 10.1007/s00330-022-09059-w. Epub 2022 Aug 27.

Abstract

Objectives: Cardiac implantable electronic device (CIED)-induced metal artefacts possibly significantly diminish the diagnostic value of magnetic resonance imaging (MRI), particularly cardiac MR (CMR). Right-sided generator implantation, wideband late-gadolinium enhancement (LGE) technique and raising the ipsilateral arm to the generator during CMR scanning may reduce the CIED-induced image artefacts. We assessed the impact of generator location and the arm-raised imaging position on the CIED-induced artefacts in CMR.

Methods: We included all clinically indicated CMRs performed on patients with normal cardiac anatomy and a permanent CIED with endocardial pacing leads between November 2011 and October 2019 in our institution (n = 171). We analysed cine and LGE sequences using the American Heart Association 17-segment model for the presence of artefacts.

Results: Right-sided generator implantation and arm-raised imaging associated with a significantly increased number of artefact-free segments. In patients with a right-sided pacemaker, the median percentage of artefact-free segments in short-axis balanced steady-state free precession LGE was 93.8% (IQR 9.4%, n = 53) compared with 78.1% (IQR 20.3%, n = 58) for left-sided pacemaker (p < 0.001). In patients with a left-sided implantable cardioverter-defibrillator, the median percentage of artefact-free segments reached 87.5% (IQR 6.3%, n = 9) using arm-raised imaging, which fell to 62.5% (IQR 34.4%, n = 9) using arm-down imaging in spoiled gradient echo short-axis cine (p = 0.02).

Conclusions: Arm-raised imaging represents a straightforward method to reduce CMR artefacts in patients with left-sided generators and can be used alongside other image quality improvement methods. Right-sided generator implantation could be considered in CIED patients requiring subsequent CMR imaging to ensure sufficient image quality.

Key points: • Cardiac implantable electronic device (CIED)-induced metal artefacts may significantly diminish the diagnostic value of an MRI, particularly in cardiac MRIs. • Raising the ipsilateral arm relative to the CIED generator is a cost-free, straightforward method to significantly reduce CIED-induced artefacts on cardiac MRIs in patients with a left-sided generator. • Right-sided generator implantation reduces artefacts compared with left-sided implantation and could be considered in CIED patients requiring subsequent cardiac MRIs to ensure adequate image quality in the future.

Keywords: Artefacts; Defibrillators, Implantable; Electrodes, Implanted; Magnetic resonance imaging; Pacemaker, Artificial.