Improved myocardial scar visualization with fast free-breathing motion-compensated black-blood T1-rho-prepared late gadolinium enhancement MRI

Diagn Interv Imaging. 2022 Dec;103(12):607-617. doi: 10.1016/j.diii.2022.07.003. Epub 2022 Aug 10.

Abstract

Purpose: Clinical guidelines recommend the use of bright-blood late gadolinium enhancement (BR-LGE) for the detection and quantification of regional myocardial fibrosis and scar. This technique, however, may suffer from poor contrast at the blood-scar interface, particularly in patients with subendocardial myocardial infarction. The purpose of this study was to assess the clinical performance of a two-dimensional black-blood LGE (BL-LGE) sequence, which combines free-breathing T1-rho-prepared single-shot acquisitions with an advanced non-rigid motion-compensated patch-based reconstruction.

Materials and methods: Extended phase graph simulations and phantom experiments were performed to investigate the performance of the motion-correction algorithm and to assess the black-blood properties of the proposed sequence. Fifty-one patients (37 men, 14 women; mean age, 55 ± 15 [SD] years; age range: 19-81 years) with known or suspected cardiac disease prospectively underwent free-breathing T1-rho-prepared BL-LGE imaging with inline non-rigid motion-compensated patch-based reconstruction at 1.5T. Conventional breath-held BR-LGE images were acquired for comparison purposes. Acquisition times were recorded. Two readers graded the image quality and relative contrasts were calculated. Presence, location, and extent of LGE were evaluated.

Results: BL-LGE images were acquired with full ventricular coverage in 115 ± 25 (SD) sec (range: 64-160 sec). Image quality was significantly higher on free-breathing BL-LGE imaging than on its breath-held BR-LGE counterpart (3.6 ± 0.7 [SD] [range: 2-4] vs. 3.9 ± 0.2 [SD] [range: 3-4]) (P <0.01) and was graded as diagnostic for 44/51 (86%) patients. The mean scar-to-myocardium and scar-to-blood relative contrasts were significantly higher on BL-LGE images (P < 0.01 for both). The extent of LGE was larger on BL-LGE (median, 5 segments [IQR: 2, 7 segments] vs. median, 4 segments [IQR: 1, 6 segments]) (P < 0.01), the method being particularly sensitive in segments with LGE involving the subendocardium or papillary muscles. In eight patients (16%), BL-LGE could ascertain or rule out a diagnosis otherwise inconclusive on BR-LGE.

Conclusion: Free-breathing T1-rho-prepared BL-LGE imaging with inline motion compensated reconstruction offers a promising diagnostic technology for the non-invasive assessment of myocardial injuries.

Keywords: Gadolinium enhancement; Heart; Magnetic resonance imaging; Motion; Myocardial infarction.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Cicatrix / diagnostic imaging
  • Cicatrix / pathology
  • Contrast Media*
  • Female
  • Gadolinium* / chemistry
  • Humans
  • Magnetic Resonance Imaging / methods
  • Male
  • Middle Aged
  • Myocardium / pathology
  • Predictive Value of Tests
  • Young Adult

Substances

  • Gadolinium
  • Contrast Media