Scanning a multi-channel spin-exchange relaxation-free atomic magnetometer with high spatial and time resolution

Opt Lett. 2022 Aug 1;47(15):3908-3911. doi: 10.1364/OL.465832.

Abstract

The emerging multi-channel spin-exchange relaxation-free (SERF) atomic magnetometer is a promising candidate for non-intrusive biomagnetism imaging. In this study, we propose a scanning 9-channel SERF magnetometer based on an acousto-optic modulator (AOM). Using the diffraction light of the AOM as the probe laser (with a low laser power of 1.7 mW), 9 channels were rapidly scanned by altering the diffraction angle. The scanning imaging scheme provides a new, to the best of our knowledge, approach for multi-channel magnetic field measurement and realizes a single-channel sensitivity of about 3 fT/Hz1/2, a spatial resolution of 0.6 mm, and a time resolution of about 2.7 ms, which is well suited for real-time extremely weak magnetic field imaging.