A novel temporary atrioventricular sequential pacing catheter-Characteristics and first-in-human application

J Card Surg. 2022 Oct;37(10):2991-2996. doi: 10.1111/jocs.16796. Epub 2022 Jul 28.

Abstract

Background: Sequential synchronized atrioventricular (AV) pacing provides enhanced electrophysiologic parameters which contribute to improved hemodynamic parameters and increased cardiac performance to subsequently confer a clinical advantage over traditional ventricular pacing. Current temporary transvenous pacemaker catheters are limited to only one electrode which paces solely the right ventricle, thus lacking the capability to provide the optimal pacing mode. A new multilead pacemaker device was developed in response to the need for improved temporary pacing through the utilization of sequential synchronized atrioventricular pacing (TAVSP). It consists of seven preformed, preshaped nitinol wires electrodes, of which four are for intra-atrial and three for intraventricular positioning and endocardial contact, respectively. Each wire carries a ball tip designed to minimize tissue trauma and provide a high current density for adequate myocardial capture. The device is not yet Food and Drug Administration approved.

Objective: To present the unique structural components and mechanical properties of a novel sequential synchronized AV pacing device for temporary insertion and to report its first-in-human application with an analysis of the early clinical experience.

Methods: Following a process of development and proof of concept of the novel pacing modality in an animal model which demonstrated feasibility and safety, a series of patients who were candidates for the device application was identified. During left and right heart catheterization, the novel temporary pacing catheter was inserted transvenously and delivered in most patients under fluoroscopy or echocardiography. The catheter was deployed to its target right heart anatomic sites and then activated in an AV sequential mode. The technical aspects, the corresponding clinical utilization, and device performance were documented and analyzed.

Results: The series included 10 enrolled subjects. During planned left and right heart catheterization, the novel TAVSP device was inserted transvenously and then delivered and deployed successfully in a timely fashion in all patients. The pacing catheter achieved proper threshold and impedance in all (100%) patients. The performance of all ventricular leads was adequate; however, in 1 (10%) patient poor performance of the atrial leads was detected. The device was successfully retrieved in all patients. No adverse arrhythmia, impaired hemodynamics, or clinical adverse events occurred. No technical difficulties, component failure, or wires thrombosis were detected. All patients sustained the device application without sequala and were discharged home.

Conclusion: Initial clinical experience with the utilization of a novel TAVSP demonstrates feasibility and safety in humans. The TAVSP modality potentially offers improved pacing capability and subsequent hemodynamic benefits over the current temporary pacing catheters. Further experience with the clinical application of this pacing catheter is warranted.

Keywords: cardiovascular pathology; cardiovascular research.