The fast calibration model for dosimetry with an electronic portal imaging device

J Appl Clin Med Phys. 2022 Nov;23(11):e13599. doi: 10.1002/acm2.13599. Epub 2022 Jul 25.

Abstract

Purpose: The aim of this study was to develop an algorithm that corrects the image of an electronic portal imaging device (EPID) of a linear accelerator so that it can be used for dosimetric purposes, such as in vivo dosimetry or quality assurance for photon radiotherapy. For that purpose, the impact of the field size, phantom thickness, and the varying spectral photon distribution within the irradiation field on the EPID image was investigated.

Methods: The EPID measurements were verified using reference measurements with ionization chambers. Therefore, absolute dose measurements with an ionization chamber and relative dose measurements with a detector array were performed. An EPID calibration and correction algorithm was developed to convert the EPID image to a dose distribution. The algorithm was validated by irradiating inhomogeneous phantoms using square fields as well as irregular IMRT fields.

Results: It was possible to correct the influence of the field size, phantom thickness on the EPID signal as well as the homogenization of the image profile by several correction factors within 0.6%. A gamma index analysis (3%, 3 mm) of IMRT fields showed a pass rate of above 99%, when comparing to the planning system.

Conclusion: The developed algorithm enables an online dose measurement with the EPID during the radiation treatment. The algorithm is characterized by a robust, non-iterative, and thus real-time capable procedure with little measuring effort and does not depend on system-specific parameters. The EPID image is corrected by multiplying three independent correction factors. Therefore, it can easily be extent by further correction factors for other influencing variables, so it can be transferred to other linear accelerators and EPID configurations.

Keywords: EPID; calibration; in vivo dosimetry; transit dosimetry.

MeSH terms

  • Calibration
  • Electronics
  • Humans
  • Particle Accelerators*
  • Phantoms, Imaging
  • Radiometry* / methods
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods