Impact of Temporary Glycemic Target Use in the Hybrid and Advanced Hybrid Closed-Loop Systems

Diabetes Technol Ther. 2022 Nov;24(11):848-852. doi: 10.1089/dia.2022.0153. Epub 2022 Aug 9.

Abstract

The Medtronic advanced hybrid closed-loop (AHCL) and MiniMed™ 670G hybrid closed-loop (HCL) systems provide the option to temporarily increase the glucose target to 150 mg/dL (8.3 mmol/L). This analysis investigated the efficacy of the AHCL compared with that of the HCL after the use of this setting. Data from 60 participants in the Fuzzy Logic Automated Insulin Regulation (FLAIR) study were used to compare the AHCL and HCL systems after the use of the temporary target (TT), and during analogous periods where this setting was not used. Differences in time in range 70-180 mg/dL between the systems were similar after the use of the TT setting and during analogous non-TT periods (interaction P = 0.87). Similar trends were observed for mean glucose, percentage time >180 mg/dL, and percentage time >250 mg/dL. Differences between AHCL and HCL systems were similar after the use of the TT setting compared with those of non-TT periods. ClinicalTrials.gov NCT03040414.

Keywords: Closed-loop systems; Continuous glucose monitoring; Insulin pump; Time in range.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Blood Glucose
  • Blood Glucose Self-Monitoring
  • Diabetes Mellitus, Type 1* / drug therapy
  • Glucose
  • Humans
  • Hypoglycemic Agents / therapeutic use
  • Insulin / therapeutic use
  • Insulin Infusion Systems*