A Comparison of a Novel Stretchable Smart Patch for Measuring Runner's Step Rates with Existing Measuring Technologies

Sensors (Basel). 2022 Jun 29;22(13):4897. doi: 10.3390/s22134897.

Abstract

A novel wearable smart patch can monitor various aspects of physical activity, including the dynamics of running, but like any new device developed for such applications, it must first be tested for validity. Here, we compare the step rate while running in place as measured by this smart patch to the corresponding values obtained utilizing ''gold standard'' MEMS accelerometers in combination with bilateral force plates equipped with HBM load cells, as well as the values provided by a three-dimensional motion capture system and the Garmin Dynamics Running Pod. The 15 healthy, physically active volunteers (age = 23 ± 3 years; body mass = 74 ± 17 kg, height = 176 ± 10 cm) completed three consecutive 20-s bouts of running in place, starting at low, followed by medium, and finally at high intensity, all self-chosen. Our major findings are that the rates of running in place provided by all four systems were valid, with the notable exception of the fast step rate as measured by the Garmin Running Pod. The lowest mean bias and LoA for these measurements at all rates were associated consistently with the smart patch.

Keywords: SINTEC; biomechanics; cadence; smart patch; stride rate; validity; wearable sensor.