Radio-frequency pulse calibration using the MISSTEC sequence

J Magn Reson. 2022 Aug:341:107260. doi: 10.1016/j.jmr.2022.107260. Epub 2022 Jun 23.

Abstract

NMR sequences are composed of multiple radio-frequency pulses. Probe adjustment, sample concentration and solvent influence the loading factor, therefore these parameters also impact the validity of flip angles. The commonly used method to calibrate RF pulses is to measure a nutation curve by varying the pulse duration. However, this method is impacted by off-resonance effects, radiation damping and B1 and B0 inhomogeneities. Furthermore, it is important to avoid partial saturation. In this work, the MISSTEC sequence is proposed for pulse calibration. This sequence takes only 8 s or 2 min for 1H or 13C calibration, respectively. High accuracy (with an error below 1%) was obtained for both nuclei. Therefore, the calibrations can be done rapidly and accurately. Furthermore, the MISSTEC measurement could be performed on each sample - in an automated way- before acquisitions, after which the calibration found could be automatically used.

Keywords: Automated calibration; MISSTEC; RF calibration; Spin Echo; Stimulated echo.