Quantitative Assessment of the Restoration of Original Anatomy after 3D Virtual Reduction of Long Bone Fractures

Diagnostics (Basel). 2022 Jun 2;12(6):1372. doi: 10.3390/diagnostics12061372.

Abstract

Background: The purpose of this study was to demonstrate the usefulness of 3D image-based virtual reduction by validating the evaluation criteria according to guidelines suggested by the AO Surgery Reference. Methods: For this experiment, 19 intact radial ORTHObones (ORTHObones radius, 3B Scientific, Germany, Hamburg) without any fractures were prepared. All ORTHObones with six cortical marking holes (three points on the distal part and three points on the proximal part) were scanned using a CT scanner twice (before/after intentional fracture of the ORTHObone). After the virtual reduction of all 19 ORTHObones, accuracy evaluations using the four criteria (length variation, apposition variation, alignment variation, Rotation Variation) suggested in the AO Surgery Reference were performed. Results: The mean (M) length variation was 0.42 mm, with 0.01 mm standard deviation (SD). The M apposition variation was 0.48 mm, with 0.40 mm SD. The M AP angulation variation (for alignment variation) was 3.24°, with 2.95° SD. The M lateral angulation variation (for alignment variation) was 0.09°, with 0.13° SD. The M angle of axial rotation was 1.27° with SD: 1.19°. Conclusions: The method of accuracy evaluation used in this study can be helpful in establishing a reliable plan.

Keywords: 3D image; 3D image accuracy; radial fractures; surgical plan; virtual reduction.