Background: The 3M™ Molecular Detection Assay 2-STEC Gene Screen (stx) method is based on gene amplification by the use of real-time loop-mediated isothermal amplification when used with the 3M Molecular Detection System for the rapid and specific detection of Shiga toxin gene (stx1 and/or stx2) from Shiga toxin-producing Escherichia coli (STEC) in enriched products. The 3M Molecular Detection Assay 2-STEC Gene Screen (stx) was approved as AOAC Performance Tested MethodSM Certificate No. 071903.
Objective: This matrix extension study evaluated the 3M Molecular Detection Assay 2-STEC Gene Screen (stx) method for detection of STECs in dried cannabis flower [>0.3% delta 9-tetrahydrocannabinol (THC)] and dried hemp flower (≤0.3% THC) at a 10 g test portion size.
Method: Testing followed procedures outlined in 3M Molecular Detection Assay 2-STEC Gene Screen (stx) product instructions and Standard Method Performance Requirement (SMPR®) for Detection of Shiga Toxin-Producing Escherichia coli in Cannabis and Cannabis Products (AOAC SMPR 2020.012). The method was evaluated at low, high, and non-inoculated levels.
Results: Results showed no statistically significant difference between the presumptive positive 3M Molecular Detection Assay 2-STEC Gene Screen (stx) results and the SMPR 2020.012 recommended cultural confirmations.
Conclusions: This study provides data that demonstrate the 3M Molecular Detection Assay 2-STEC Gene Screen (stx) is a reliable method for the rapid and specific detection of STECs in dried cannabis flower and dried hemp flower.
Highlights: The 3M Molecular Detection Assay 2-STEC Gene Screen (stx) method is suitable for the rapid and specific detection of STECs in dried cannabis flower and dried hemp flower.
© The Author(s) 2022. Published by Oxford University Press on behalf of AOAC INTERNATIONAL.