Highly sensitive pregnancy test kit via oriented antibody conjugation on brush-type ligand-coated quantum beads

Biosens Bioelectron. 2022 Oct 1:213:114441. doi: 10.1016/j.bios.2022.114441. Epub 2022 Jun 3.

Abstract

Lateral flow assays (LFA) enable development of portable and rapid diagnostic kits; however, their capacity to detect low levels of disease markers remains poor. Here, we report a highly sensitive pregnancy test kit as a proof of concept, by combining brush-type ligand-coated quantum beads (B-type QBs) and nanobody, which can control the antibody orientation and enhance sensitivity. The brush-type ligand provided excellent dispersion stability and high-binding capacity toward antibody. Fc-binding nanobody increased the antigen-binding capacity of conjugated antibodies on the B-type QBs. To facilitate convenient acquisition of the LFA results, we developed a smartphone-based reader with a 3D-printed optical imaging module, and validated the diagnostic performance of the sensing platform. The pregnancy test kit achieved a 5.1 pg mL-1 limit of detection, corresponding to the levels for early-stage detection of heart disease and malaria. Our LFA application can potentially be expanded to diagnosis other diseases by simply changing the antibody pair in the kit.

Keywords: Android application; Antibody orientation; Brush-type ligand; Portable detector; Pregnancy test kit; Quantum bead.

MeSH terms

  • Antibodies
  • Biosensing Techniques* / methods
  • Female
  • Humans
  • Ligands
  • Pregnancy
  • Pregnancy Tests*

Substances

  • Antibodies
  • Ligands