Cochlear base length as predictor for angular insertion depth in incomplete partition type 2 malformations

Int J Pediatr Otorhinolaryngol. 2022 Aug:159:111204. doi: 10.1016/j.ijporl.2022.111204. Epub 2022 Jun 9.

Abstract

Introduction: The preoperative determination of suitable electrode array lengths for cochlear implantation in inner ear malformations is a matter of debate. The choice is usually based on individual experience and the use of intraoperative probe electrodes. The purpose of this case series was to evaluate the applicability and precision of an angular insertion depth (AID) prediction method, based on a single measurement of the cochlear base length (CBL).

Methods: We retrospectively measured the CBL in preoperative computed tomography (CT) images in 10 ears (8 patients) with incomplete partition type 2 malformation. With the known electrode length (linear insertion depth, LID) the AID at full insertion was retrospectively predicted for each ear with a heuristic equation derived from non-malformed cochleae. Using the intra- or post-implantation cone beam CT images, the actual AID was assessed and compared. The deviations of the predicted from the actual insertion angles were quantified (clinical prediction error) to assess the precision of this single-measure estimation.

Results: Electrode arrays with 15 mm (n = 3), 19 mm (n = 2), 24 mm (n = 3), and 26 mm (n = 2) length were implanted. Postoperative AIDs ranged from 211° to 625°. Clinical AID prediction errors from -64° to 62° were observed with a mean of 0° (SD of 44°). In two ears with partial insertion of the electrode, the predicted AID was overestimated. The probe electrode was intraoperatively used in 9/10 cases.

Conclusion: The analyzed method provides good predictions of the AID based on LID and CBL. It does not account for incomplete insertions, which lead to an overestimation of the AID. The probe electrode is useful and well established in clinical practice. The investigated method could be used for patient-specific electrode length selection in future patients.

Keywords: A-value; Cochlear duct length; Incomplete partition type 2; Insertion depth; Mondini dysplasia.

MeSH terms

  • Cochlea / diagnostic imaging
  • Cochlea / surgery
  • Cochlear Implantation* / methods
  • Cochlear Implants*
  • Electrodes, Implanted
  • Humans
  • Retrospective Studies
  • Tomography, X-Ray Computed / methods