Evaluation of a self-imaging SD-OCT system designed for remote home monitoring

BMC Ophthalmol. 2022 Jun 10;22(1):261. doi: 10.1186/s12886-022-02458-z.

Abstract

Purpose: To compare identification rates of retinal fluid of the Notal Vision Home Optical Coherence Tomography (OCT) device (NVHO) when used by people with age-related macular degeneration (AMD) to those captured by a commercial OCT.

Methods: Prospective, cross-sectional study where patients underwent commercial OCT imaging followed by self-imaging with either the NVHO 2.5 or the NVHO 3 in clinic setting. Outcomes included patients' ability to acquire analyzable OCT images with the NVHO and to compare those with commercial images.

Results: Successful images were acquired with the NVHO 2.5 in 469/531 eyes (88%) in 264/290 subjects (91%) with the mean (SD) age of 78.8 (8.8); 153 (58%) were female with median visual acuity (VA) of 20/40. In the NVHO 3 cohort, 69 eyes of 45 subjects (93%) completed the self-imaging. Higher rates of successful imaging were found in eyes with VA ≥ 20/320. Positive percent agreement/negative percent agreement for detecting the presence of subretinal and/or intraretinal fluid when reviewing for fluid in three repeated volume scans were 97%/95%, respectively for the NVHO v3.

Conclusion: Self-testing with the NVHO can produce high quality images suitable for fluid identification by human graders, suggesting the device may be able to complement standard-of-care clinical assessments and treatments.

Keywords: Age-related macular degeneration; HOME OCT; Home monitoring.