Frequency-Domain Detection for Frequency-Division Multiplexing QEPAS

Sensors (Basel). 2022 May 26;22(11):4030. doi: 10.3390/s22114030.

Abstract

To achieve multi-gas measurements of quartz-enhanced photoacoustic spectroscopy (QEPAS) sensors under a frequency-division multiplexing mode with a narrow modulation frequency interval, we report a frequency-domain detection method. A CH4 absorption line at 1653.72 nm and a CO2 absorption line at 2004.02 nm were investigated in this experiment. A modulation frequency interval of as narrow as 0.6 Hz for CH4 and CO2 detection was achieved. Frequency-domain 2f signals were obtained with a resolution of 0.125 Hz using a real-time frequency analyzer. With the multiple linear regressions of the frequency-domain 2f signals of various gas mixtures, small deviations within 2.5% and good linear relationships for gas detection were observed under the frequency-division multiplexing mode. Detection limits of 0.6 ppm for CH4 and 2.9 ppm for CO2 were simultaneously obtained. With the 0.6-Hz interval, the amplitudes of QEPAS signals will increase substantially since the modulation frequencies are closer to the resonant frequency of a QTF. Furthermore, the frequency-domain detection method with a narrow interval can realize precise gas measurements of more species with more lasers operating under the frequency-division multiplexing mode. Additionally, this method, with a narrow interval of modulation frequencies, can also realize frequency-division multiplexing detection for QEPAS sensors under low pressure despite the ultra-narrow bandwidth of the QTF.

Keywords: frequency-division multiplexing; frequency-domain signal; photoacoustic spectroscopy; quartz tuning fork.