Inlay versus onlay humeral design for reverse shoulder arthroplasty: a systematic review and meta-analysis

J Shoulder Elbow Surg. 2022 Nov;31(11):2410-2420. doi: 10.1016/j.jse.2022.05.002. Epub 2022 Jun 4.

Abstract

Background: Since the introduction of the Grammont-style reverse total shoulder arthroplasty, the humeral stem design has been modified with improved clinical outcomes. Two distinct humeral designs have been used extensively: the inlay design, in which the humeral tray is seated within the metaphysis, and the onlay design, in which the humeral tray sits on the metaphysis at the level of the humeral neck cut. The purpose of this systematic review was to determine whether there are differences in clinical outcomes and complication rates between these designs.

Methods: The Preferred Reporting Items for Systematic Reviews and Meta-analyses (PRISMA) guidelines were used to perform this systematic review. A search of MEDLINE, PubMed, and Embase was performed to identify all studies comparing the clinical results of both humeral designs. Primary outcomes included patient-reported outcome measures, shoulder range of motion, and incidence of complications.

Results: From the 156 identified publications, 12 studies were included in the final review. A total of 1447 patients were included, with a minimum follow-up period of 12 months. At final follow-up, both implants demonstrated significant improvements in comparison to preoperative baseline. On comparison of the inlay vs. onlay groups, the American Shoulder and Elbow Surgeons score was higher in the inlay group (mean difference, 2.53 [95% confidence interval, 0.27-4.78]; P = .03). Postoperative motion, even if statistically greater in the onlay group (differences of 5° in forward flexion [P < .001], 3° in abduction [P = .003], and 4° in external rotation [P < .001]), was not clinically different. On comparison of complications, the inlay group showed more instances of scapular notching (93 of 322 patients vs. 70 of 415 patients; odds ratio, 0.35; P < .001) but fewer scapular spine fractures (26 of 727 patients vs. 21 of 559 patients, P = .09).

Discussion: Inlay and onlay humeral tray designs in reverse total shoulder arthroplasty demonstrate similar clinical improvements postoperatively. Onlay implants have a low rate of scapular notching but a higher rate of scapular spine fracture. Understanding the strengths and weaknesses of the 2 humeral tray designs is important to provide surgeons with options to tailor surgical plans for high-risk patients.

Keywords: Humeral tray design; RTSA; inlay; onlay; scapular notching; scapular spine fracture.

Publication types

  • Meta-Analysis
  • Systematic Review
  • Review

MeSH terms

  • Arthroplasty, Replacement, Shoulder* / adverse effects
  • Humans
  • Humerus / surgery
  • Range of Motion, Articular
  • Scapula / surgery
  • Shoulder Joint* / surgery