Manufacturer Signal-to-Cutoff Threshold Underestimates Cumulative Incidence of SARS-CoV-2 Infection: Evidence from the Los Angeles Firefighters Study

J Appl Lab Med. 2022 Sep 1;7(5):1169-1174. doi: 10.1093/jalm/jfac034.

Abstract

Background: The objective of this analysis was to compare the performance sensitivity and specificity of manufacturer-recommended signal-to-cutoff (S/Co) thresholds with modified S/Co values to estimate the prevalence of SARS-CoV-2-specific antibodies in a cohort of firefighters with a known infection history.

Methods: Plasma venipuncture samples were used for serologic analysis of firefighters in Los Angeles, CA, USA, in October 2020. Seropositivity was assessed using the manufacturer's recommended S/Co (≥1.4 IgG) and modified S/Co thresholds based on measured antibody levels in 178 negative control patients who had blood drawn prior to the emergence of COVID-19. Optimal S/Co threshold was determined by receiver operating characteristic (ROC) curve analysis.

Results: Of 585 firefighters included in the study, 52 (8.9%) reported having a PCR-positive test history prior to antibody testing. Thirty-five (67.3%) firefighters with a previous PCR-positive test were seropositive based on the manufacturer S/Co thresholds, consistent with an estimated 67.3% sensitivity and 100% specificity. After evaluating multiple modified S/Co thresholds based on pre-pandemic negative samples, a modified S/Co of 0.36 was found to yield optimal sensitivity (88.5%) and specificity (99.4%) by ROC curve analysis. This modified threshold improved serostatus classification accuracy by 21.2%.

Conclusions: S/Co thresholds based on known negative samples significantly increase seropositivity and more accurately estimate cumulative incidence of disease compared to manufacturer-based thresholds.

MeSH terms

  • Antibodies, Viral
  • COVID-19* / diagnosis
  • COVID-19* / epidemiology
  • Firefighters*
  • Humans
  • Incidence
  • Los Angeles / epidemiology
  • SARS-CoV-2

Substances

  • Antibodies, Viral