Feasibility assessment of patient-controlled EEG home-monitoring: More results from the HOMEONE study

Clin Neurophysiol. 2022 Aug:140:12-20. doi: 10.1016/j.clinph.2022.04.021. Epub 2022 May 11.

Abstract

Objective: The feasibility phase of the HOME (Home-Monitoing and Education) project aims to show the practical feasibility of Electroencephalography (EEG)home-monitoring using a patient-controlled mobile system. Its objective is to assess the potential diagnostic and therapeutic yields of home-monitoring compared to conventional healthcare.

Methods: 16 office-based practitioners chose 97 patients and recorded standard 20-minute EEGs using conventional recorders. After training, the same patients used a patient-controlled mobile dry electrode EEG system in their home environment. The practitioners in charge and two additional raters assessed all recordings. We conducted inter-rater and intra-rater comparisons between the diagnostic findings.

Results: 89 patients successfully conducted home-monitoring recordings. The intra-rater comparison results for the diagnostic findings of the conventional recordings and the patient-made recordings show a fair Cohen's kappa value (0.21). Additionally, we documented a change of patient management in 9 cases.

Conclusions: The feasibility of EEG home-monitoring using a patient-controlled device is confirmed. The yield of EEG home-monitoring comprises information that can influence patient management.

Significance: Patient-controlled EEG home-monitoring is feasible as part of routine care for neurological outpatients as its technical efficacy and practical feasibility are shown and significantly positive effects on patient management are evidenced.

Keywords: Change of management; EEG home-monitoring; Feasibility assessment; Intra-individual comparison; Tele-EEG.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electrodes
  • Electroencephalography* / methods
  • Feasibility Studies
  • Humans
  • Monitoring, Physiologic