Objective We aimed to study the "in vitro" pullout strength of SpineGuard/Zavation Dynamic Surgical Guidance Z-Direct Screw (DSG Screw, SpineGuard Inc, Boulder, Colorado, USA), a screw designed to be inserted using a direct insertion technique. Methods Dynamic Surgical Guidance Screws of 5.5 and 6.5 mm were introduced into polyurethane blocks with a density of 10 PCF (0,16g/cm 3 ). According to the experimental group, screws were inserted without pilot hole, with pilot without tapping, undertapping and line-to-line tapping. Screw pullout tests were performed using a universal test machine after screw insertion into polyurethane blocks. Results Screws inserted directly into the polyurethane blocks without pilot hole and tapping showed a statistically higher pullout strength. Insertion of the screw without tapping or with undertapping increases the pullout screw strength compared with line-to-line tapping. Conclusion Dynamic Surgical Guidance Screw showed the highest pullout strength after its insertion without pilot hole and tapping.
Keywords: biomechanical phenomena/physiology; bone screws; pedicle screw; spinal fusion.
Sociedade Brasileira de Ortopedia e Traumatologia. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commecial purposes, or adapted, remixed, transformed or built upon. ( https://creativecommons.org/licenses/by-nc-nd/4.0/ ).