Capsule-Based dry powder inhaler evaluation using CFD-DEM simulations and next generation impactor data

Eur J Pharm Sci. 2022 Aug 1:175:106226. doi: 10.1016/j.ejps.2022.106226. Epub 2022 May 25.

Abstract

Capsule-based, single-dose dry powder inhalers (DPIs) are commonly-used devices to deliver medications to the lungs. This work evaluates the effect of the drug/excipient adhesive bonding and the DPI resistances on the aerosol performance using a combination of empirical multi-stage impactor data and a fully-coupled computational fluid dynamics (CFD) and discrete element method (DEM) model. Model-predicted quantities show that the primary modes of powder dispersion are a function of the device resistance. Lowering the device resistance increases its capacity to transport a wider range of particle size classes toward the outlet and generate more intense turbulence upstream therein. On the other hand, a higher device resistance increases the velocity of the tangential airflow along the device walls, which in turn increases the intensity of particle/device impaction. Correlating model data and experimental results shows that these differing powder dispersion mechanisms affect different formulations differently, with finer aerosols tending to result when pairing a lower resistance device with formulations that exhibit low API/excipient adhesion, or when pairing a high resistance device with more cohesive formulations.

Keywords: CFD-DEM; Computational fluid dynamics; Discrete element method; Dry powder inhalers; Inhalation drug delivery.

MeSH terms

  • Administration, Inhalation
  • Aerosols
  • Dry Powder Inhalers*
  • Equipment Design
  • Excipients
  • Hydrodynamics*
  • Particle Size
  • Powders

Substances

  • Aerosols
  • Excipients
  • Powders