Abstract

Objectives: Sutureless aortic valve replacement (SU-SAVR) has been associated with higher rates of permanent pacemaker (PPM) compared with conventionally implanted aortic bioprostheses. The purpose of this study was to determine the incidence, predictors and mid-term prognostic impact of PPM after Perceval (Livanova, London, UK) SU-SAVR in low-risk patients.

Methods: A total of 400 consecutive low-risk (EuroSCORE II < 4%) patients without prior pacemaker who underwent surgical aortic valve replacement with the Perceval prosthesis from 2013 to 2019 in 2 centres were included. Baseline, clinical and electrocardiographic parameters, procedural characteristics and follow-up data were collected.

Results: PPM was required in 36 (9%) patients after SU-SAVR, with a median time between the procedure and PPM implantation of 7.5 (4.5-10.5) days. Older age and prior right bundle branch block (RBBB) were associated with an increased risk of PPM (P < 0.05 for all), but only baseline RBBB was found to be an independent predictor of new PPM requirement (odds ratio: 2.60, 95% confidence interval: 1.15-5.81; P = 0.022). At a median follow-up of 3.4 (2.3-4.5) years, there were no differences between groups in mortality (PPM: 36%, no PPM: 22%, P = 0.105) or heart failure rehospitalization (PPM: 25%, no PPM: 21%, P = 0.839).

Conclusions: About 1 out of 10 low-risk patients with aortic stenosis undergoing SU-SAVR with the Perceval prosthesis required PPM implantation. Prior RBBB determined an increased risk (close to 3-fold) of PPM following the procedure. PPM was not associated with a higher risk of clinical events at 3-year follow-up.

Keywords: Aortic valve stenosis; Cardiac pacemaker; Heart conduction system.

MeSH terms

  • Aortic Valve / surgery
  • Aortic Valve Stenosis*
  • Bundle-Branch Block / epidemiology
  • Heart Valve Prosthesis Implantation* / methods
  • Heart Valve Prosthesis* / adverse effects
  • Humans
  • Incidence
  • Pacemaker, Artificial* / adverse effects
  • Prognosis
  • Risk Factors
  • Transcatheter Aortic Valve Replacement* / adverse effects
  • Treatment Outcome