Heavy Metal Enhancement Technique for Diaminobenzidine in Immunohistochemistry Enables Ultrastructural Observation by Low-vacuum Scanning Electron Microscopy

J Histochem Cytochem. 2022 Jun;70(6):427-436. doi: 10.1369/00221554221102996. Epub 2022 May 25.

Abstract

Low-vacuum scanning electron microscopy (LV-SEM) is a powerful tool that allows to observe light microscopic specimens with periodic acid-silver methenamine (PAM) staining at a higher magnification, simply by removing the coverslip. However, it is not suitable for observation of immunohistochemistry (IHC) using 3,3'-diaminobenzidine (DAB) due to insufficient backscattered electron image. Traditional heavy metal enhancement techniques for DAB in IHC, (1) osmium tetroxide and iron, (2) cobalt, (3) methenamine silver (Ag), (4) gold chloride (Gold), and (5) both Ag and Gold (Ag + Gold), were examined by LV-SEM. Tissue specimens from Thy1.1 glomerulonephritis rat kidney stained with α-smooth muscle actin and visualized with DAB were enhanced by each of these enhancement methods. We found, in light microscopic and LV-SEM, that the enhancement with Ag, Gold, or Ag + Gold had better intensity and contrast than others. At a higher magnification, Ag + Gold enhancement showed high intensity and low background, although only Ag or Gold enhancement had nonspecific background. Even after observation by LV-SEM, the quality of specimens was maintained after remounting the coverslip. It was also confirmed that Ag + Gold enhancement could be useful for IHC using clinical human renal biopsy. These findings indicate that Ag + Gold provided an adequate enhancement in IHC for both LM and LV SEM observation.

Keywords: heavy metal enhancement; immunohistochemistry; low-vacuum SEM; renal pathology.

MeSH terms

  • Animals
  • Gold*
  • Immunohistochemistry
  • Microscopy, Electron, Scanning
  • Osmium Tetroxide*
  • Rats
  • Vacuum

Substances

  • Gold
  • Osmium Tetroxide