Nanocoral-like Polyaniline-Modified Graphene-Based Electrochemical Paper-Based Analytical Device for a Portable Electrochemical Sensor for Xylazine Detection

ACS Omega. 2022 Apr 12;7(16):13913-13924. doi: 10.1021/acsomega.2c00295. eCollection 2022 Apr 26.

Abstract

A portable electrochemical device for xylazine detection is presented for the first time. An electrochemical paper-based analytical device (ePAD) was integrated with a smartphone. The fabrication of the ePAD involved wax printing, low-tack transfer tape, and cutting and screen-printing techniques. Graphene ink was coated on the substrate and modified with nanocoral-like polyaniline, providing an electron transfer medium with a larger effective surface area that promoted charge transfer. The conductive ink on the ePAD presented a thickness of 25.0 ± 0.9 μm for an effective surface area of 0.374 cm2. This sensor was then tested directly on xylazine using differential pulse voltammetry. Two linear responses were obtained: from 0.2 to 5 μg mL-1 and from 5 to 100 μg mL-1. The detection limit was 0.06 μg mL-1. Reproducibility was tested on 10 preparations. The relative standard deviation was less than 5%. The applicability of the sensor was evaluated with beverage samples spiked with trace xylazine. Recoveries ranged from 84 ± 4 to 105 ± 2%. The developed sensor demonstrated excellent accuracy in the detection of trace xylazine. It would be possible to develop the portable system to detect various illicit drugs to aid forensic investigations.