Validation of aortic valve pressure gradient quantification using semi-automated 4D flow CMR pipeline

BMC Res Notes. 2022 Apr 29;15(1):151. doi: 10.1186/s13104-022-06033-z.

Abstract

Objective: Doppler echocardiographic aortic valve peak velocity and peak pressure gradient assessment across the aortic valve (AV) is the mainstay for diagnosing aortic stenosis. Four-dimensional flow cardiovascular magnetic resonance (4D flow CMR) is emerging as a valuable diagnostic tool for estimating the peak pressure drop across the aortic valve, but assessment remains cumbersome. We aimed to validate a novel semi-automated pipeline 4D flow CMR method of assessing peak aortic value pressure gradient (AVPG) using the commercially available software solution, CAAS MR Solutions, against invasive angiographic methods.

Results: We enrolled 11 patients with severe AS on echocardiography from the EurValve programme. All patients had pre-intervention doppler echocardiography, invasive cardiac catheterisation with peak pressure drop assessment across the AV and 4D flow CMR. The peak AVPG was 51.9 ± 35.2 mmHg using the invasive pressure drop method and 52.2 ± 29.2 mmHg for the 4D flow CMR method (semi-automated pipeline), with good correlation between the two methods (r = 0.70, p = 0.017). Assessment of AVPG by 4D flow CMR using the novel semi-automated pipeline method shows excellent agreement to invasive assessment when compared to doppler-based methods and advocate for its use as complementary to echocardiography.

Keywords: Aortic stenosis; Cardiac MRI; Validation.

MeSH terms

  • Aortic Valve Stenosis* / diagnostic imaging
  • Aortic Valve Stenosis* / pathology
  • Aortic Valve* / diagnostic imaging
  • Blood Flow Velocity
  • Echocardiography
  • Echocardiography, Doppler
  • Humans