The influence of non-conformal grid interfaces on the results of large eddy simulation of centrifugal blood pumps

Artif Organs. 2022 Sep;46(9):1804-1816. doi: 10.1111/aor.14263. Epub 2022 Apr 26.

Abstract

Background: Computational fluid dynamics has been widely used to assist the design and evaluation of blood pumps. Discretization errors associated with computational grid may influence the credibility of numerical simulations. Non-conformal grid interfaces commonly exist in rotary machines, including rotary blood pumps. Should grid size across the interface differ greatly, large errors may occur.

Methods: This study explored the effects of non-conformal grid interface on the prediction of the flow field and hemolysis in blood pumps using large eddy simulation (LES). Two benchmarks, a nozzle model and a centrifugal blood pump were chosen as test cases.

Results: This study found that non-conformal grid interfaces with considerable change of grid sizes led to discontinuities of flow variables and brought errors to metrics such as pressure head (7%) and hemolysis (up to 14%).

Conclusions: The results on the full unstructured grid are more accurate with negligible changes of flow variables across the non-conformal grid interface. A full unstructured grid should be employed for centrifugal blood pumps to minimize the influence of non-conformal grid interfaces for LES simulations.

Keywords: blood pump; computational grid; flow field; large eddy simulation; non-conformal grid interface.

MeSH terms

  • Computer Simulation
  • Heart-Assist Devices* / adverse effects
  • Hemolysis*
  • Humans
  • Hydrodynamics