Efficient measurement of airborne viable viruses using the growth-based virus aerosol concentrator with high flow velocities

J Hazard Mater. 2022 Jul 15:434:128873. doi: 10.1016/j.jhazmat.2022.128873. Epub 2022 Apr 6.

Abstract

Growth tube collectors (GTCs) are used to sample virus aerosols because of their superior viable virus recovery among air samplers. However, a major limitation of such samplers is that they operate at low flow rates compared to many inertia-based air samplers. Herein, we demonstrated efficient measurements of airborne MS2 and T3 viruses using a GTC that can implement high flow velocities for higher flow rates per tube, which we refer to as the growth-based virus aerosol concentrator (GVC), via qPCR and the plaque assay technique. The GVC exhibited a flow rate of up to 6 L/min, where the average sampling flow velocity was 5.09 m/s, 22 times higher than those used in the GTCs, for a single tube with a diameter of 5 mm. The count median diameter of the size-increased particles at the exit of the initiator was measured to be 1.44 µm at 6 L/min, considerably smaller than those observed in conventional GTCs. Nevertheless, the measurement of airborne MS2 and T3 viruses using the GVC showed a high concentration (high enrichment ratio of 109,458 at 10-min sampling) of viruses in a sampling medium, with a high viable virus percentage (> 90%) and physical collection efficiency (> 90%) at 6 L/min, which shows the potential for rapid on-site detection of airborne viruses.

Keywords: Air sampler; Airborne virus measurement; Laminar flow water condensation; Virus aerosols.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aerosols / analysis
  • Air Microbiology*
  • Environmental Monitoring / methods
  • Particle Size
  • Viruses*

Substances

  • Aerosols