Single center experience of transcatheter closure of mitral and aortic Paravalvular leaks using the new rechristened rectangular Amplatzer PVL plug

Indian Heart J. 2022 May-Jun;74(3):239-241. doi: 10.1016/j.ihj.2022.04.004. Epub 2022 Apr 12.

Abstract

Amplatzer Paravalvular leak (PVL) plug is rectangular in shape, that might suit closure of crescentic PVL. Among 79 transcatheter PVL closures from a single center, a subgroup of 16 patients who received Amplatzer PVL plugs were analyzed. All procedures were successful, as the plug auto-oriented to the leak, without mechanical leaflet interference, though needing additional 31 devices. Two patients needed an elective re-intervention. NYHA class improved from III-IV before procedure to less than II after procedure. Complications were not directly related to the Amplatzer PVL plug. The rectangular shape for crescentic leaks, auto-orientation and non-interference with mechanical leaflets were good attributes.

Keywords: Crescentic shape; Oblong plug; Periprosthetic leak.

MeSH terms

  • Cardiac Catheterization / methods
  • Heart Valve Prosthesis Implantation* / methods
  • Heart Valve Prosthesis* / adverse effects
  • Humans
  • Prosthesis Failure
  • Treatment Outcome