Handheld laser scanning microscope catheter for real-time and in vivo confocal microscopy using a high definition high frame rate Lissajous MEMS mirror

Biomed Opt Express. 2022 Feb 15;13(3):1497-1505. doi: 10.1364/BOE.447558. eCollection 2022 Mar 1.

Abstract

A handheld confocal microscope using a rapid MEMS scanning mirror facilitates real-time optical biopsy for simple cancer diagnosis. Here we report a handheld confocal microscope catheter using high definition and high frame rate (HDHF) Lissajous scanning MEMS mirror. The broad resonant frequency region of the fast axis on the MEMS mirror with a low Q-factor facilitates the flexible selection of scanning frequencies. HDHF Lissajous scanning was achieved by selecting the scanning frequencies with high greatest common divisor (GCD) and high total lobe number. The MEMS mirror was fully packaged into a handheld configuration, which was coupled to a home-built confocal imaging system. The confocal microscope catheter allows fluorescence imaging of in vivo and ex vivo mouse tissues with 30 Hz frame rate and 95.4% fill factor at 256 × 256 pixels image, where the lateral resolution is 4.35 μm and the field-of-view (FOV) is 330 μm × 330 μm. This compact confocal microscope can provide diverse handheld microscopic applications for real-time, on-demand, and in vivo optical biopsy.