Frequency-Based Maternal Electrocardiogram Attenuation for Fetal Electrocardiogram Analysis

Ann Biomed Eng. 2022 Jul;50(7):836-846. doi: 10.1007/s10439-022-02959-4. Epub 2022 Apr 11.

Abstract

Fetal electrocardiogram (ECG) waveform analysis along with cardiac time intervals (CTIs) measurements are critical for the management of high-risk pregnancies. Currently, there is no system that can consistently and accurately measure fetal ECG. In this work, we present a new automatic approach to attenuate the maternal ECG in the frequency domain and enhance it with measurable CTIs. First, the coherent components between the maternal ECG and abdominal ECG were identified and subtracted from the latter in the frequency domain. The residual was then converted into the time domain using the inverse Fourier transform to yield the fetal ECG. This process was improved by averaging multiple beats. Two fetal cardiologists, blinded to the method, assessed the quality of fetal ECG based on a grading system and measured the CTIs. We evaluated the fetal ECG quality of our method and time-based methods using one synthetic dataset, one human dataset available in the public domain, and 37 clinical datasets. Among the 37 datasets analyzed, the mean average (± standard deviation) grade was 3.49 ± 1.22 for our method vs. 2.64 ± 1.26 for adaptive interference cancellation (p-value < 0.001), thus showing the frequency-based fetal ECG extraction was the superior method, as assessed from our clinicians' perspectives. This method has the potential for use in clinical settings.

Keywords: Abdominal electrocardiogram; Fetal electrocardiogram; Maternal electrocardiogram; Spectral coherence.

MeSH terms

  • Abdomen
  • Algorithms
  • Electrocardiography* / methods
  • Female
  • Fetus*
  • Heart
  • Humans
  • Pregnancy
  • Signal Processing, Computer-Assisted