[A Method for Evaluating the T2-weighting Effect in MRI]

Nihon Hoshasen Gijutsu Gakkai Zasshi. 2022 Apr 20;78(4):357-363. doi: 10.6009/jjrt.2022-1189. Epub 2022 Mar 15.
[Article in Japanese]

Abstract

Purpose: To propose a method for evaluating the T2*-weighting effect in MRI.

Methods: Multiple solutions with different concentrations of a superparamagnetic iron oxide contrast agent were made and their signal intensities on T2*-weighted images were measured. The relationship between iron concentration and signal intensity was determined, and we simulated an iron concentration map representing a simplified model of a brain microbleed and converted the pixel values in the map to signal intensity based on the determined relationship, generating a simulated T2*-weighted image. An 'S-value' parameter was defined to evaluate the low-intensity regions in the simulated image. S-values were obtained using T2*-weighted sequences acquired with different echo time (TE) values on three MRI scanners (Philips 1.5 T, GE 3.0 T, and Siemens 3.0 T). Another parameter (A-value) defined by the American Society for Testing and Materials (ASTM-F2119) for assessing artifacts was applied to evaluate the weighting effect in the T2*-weighted image of a laboratory-made susceptibility-effect phantom.

Results: With all three scanners, the S-values increased as the TE increased, indicating enhancement of the T2*-weighting effect. For every TE, the S-values obtained for the Philips scanner were the largest, followed by those for the GE and Siemens scanners. The results of this comparative evaluation were similar to those obtained using A-values.

Conclusion: Comparisons with the established A-value parameter showed our proposed method for the quantitative evaluation of the T2*-weighting effect using S-values to be valid. The proposed method has the advantage that the S-values do not depend on a specific susceptibility-effect phantom.

Keywords: T2*-weighted image; magnetic resonance imaging (MRI); superparamagnetic iron oxide (SPIO); susceptibility effect.

MeSH terms

  • Brain / diagnostic imaging
  • Contrast Media*
  • Iron
  • Magnetic Resonance Imaging* / methods
  • Phantoms, Imaging

Substances

  • Contrast Media
  • Iron