On-Orbit Absolute Radiometric Calibration and Validation of ZY3-02 Satellite Multispectral Sensor

Sensors (Basel). 2022 Mar 7;22(5):2066. doi: 10.3390/s22052066.

Abstract

This study described the on-orbit vicarious radiometric calibration of Chinese civilian high-resolution stereo mapping satellite ZY3-02 multispectral imager (MUX). The calibration was based on gray-scale permanent artificial targets, and multiple radiometric calibration tarpaulins (tarps) using a reflectance-based approach between July and September 2016 at Baotou calibration site in China was described. The calibration results reveal a good linear relationship between DN and TOA radiances of ZY3-02 MUX. The uncertainty of this radiometric calibration was 4.33%, indicating that radiometric coefficients of ZY3-02 MUX are reliable. A detailed discussion on the validation analysis of the comparison results between the different radiometric calibration coefficients is presented in this paper. To further validate the reliability of the three coefficients, the calibrated ZY3-02 MUX was compared with Landsat-8 Operational Land Imager (OLI). The results also indicate that radiometric characteristics of ZY3-02 MUX imagery are reliable and highly accurate for quantitative applications.

Keywords: TOA radiance; ZY3-02 satellite; radiometric calibration; reflectance-based approach; the Baotou site.