Beamline commissioning for microscopic measurements with ultraviolet and soft X-ray beam at the upgraded beamline BL-13B of the Photon Factory

J Synchrotron Radiat. 2022 Mar 1;29(Pt 2):400-408. doi: 10.1107/S160057752200090X. Epub 2022 Feb 16.

Abstract

Beamline 13 of the Photon Factory has been in operation since 2010 as a vacuum ultraviolet and soft X-ray undulator beamline for X-ray photoelectron spectroscopy (XPS), X-ray absorption spectroscopy (XAS), and angle-resolved photoelectron spectroscopy (ARPES) experiments. The beamline and the end-station at branch B have been recently upgraded, enabling microscopic XPS, XAS, and ARPES measurements to be performed. In 2015, a planar undulator insertion device was replaced with an APPLE-II (advanced planar polarized light emitter II) undulator. This replacement allows use of linear, circular, and elliptical polarized light between 48 and 2000 eV with photon intensities of 109-1013 photons s-1. For microscopic measurements, a toroidal post-mirror was renewed to have more focused beam with profile sizes of 78 µm (horizontal) × 15 µm (vertical) and 84 µm × 11 µm at photon energies of 100 and 400 eV, respectively. A high-precision sample manipulator composed of an XYZ translator, a rotary feedthrough, and a newly developed goniometer, which is essential for microscopic measurements, has been used to control a sample specimen in six degrees of freedom, i.e. translation in the X, Y, and Z directions and rotation in the polar, azimuthal, and tilt directions. To demonstrate the performance of the focused beams, one- and two-dimensional XPS and XAS scan measurements of a copper grid have been performed. It was indicated from analysis of XPS and XAS intensity maps that the actual spatial resolution can be determined by the beam size.

Keywords: VUV-SX beamline; X-ray absorption spectroscopy; X-ray photoelectron spectroscopy; angle-resolved photoelectron spectroscopy; microscopic measurement.