Feasibility of a method for low contrast CT image quality assessment using difference detail curves for abdominal scans

Z Med Phys. 2022 May;32(2):209-217. doi: 10.1016/j.zemedi.2022.01.001. Epub 2022 Feb 17.
[Article in German]

Abstract

This work describes a measurement method for assessing dose-related image-quality of CT scans based on the difference detail curve (DDC) method, and showcases its use in a low contrast setting. The method is based on a phantom consisting of elliptical slices of different sizes into which contrast object modules can be inserted. These modules contain contrast objects based on (synthetic) resin mixtures with sucrose (native) or sodium iodine (contrast medium). Mixing ratios are provided to achieve a range of clinically relevant CT-numbers with these materials. The phantom is characterized in terms of contrast accuracy, energy dependency and long-term drift with satisfying results. Contrast accuracy and energy dependency are similar to that of water or soft tissue. Image quality of 655 scans of the phantom acquired at 30 different clinical institutions and with 16 different CT scanner models from 4 manufacturers was assessed by calculating a difference detail curve (DDC) from evaluation of up to 5 human observers using a custom-made software (RadiVates) described in this work. Based on these measurements, inter-observer variability was quantified using a bootstrap method and was shown to be a large contributor to the overall variability. This work demonstrates that assessment of CT image quality is feasible with the aforementioned phantom and DDC method.

Keywords: CT image quality; CT phantom; Difference detail curve; Low contrast CT.

MeSH terms

  • Feasibility Studies
  • Humans
  • Phantoms, Imaging
  • Radiation Dosage
  • Tomography Scanners, X-Ray Computed
  • Tomography, X-Ray Computed* / methods