Non-Invasive Blood Flow Speed Measurement Using Optics

Sensors (Basel). 2022 Jan 25;22(3):897. doi: 10.3390/s22030897.

Abstract

Non-invasive measurement of the arterial blood speed gives important health information such as cardio output and blood supplies to vital organs. The magnitude and change in arterial blood speed are key indicators of the health conditions and development and progression of diseases. We demonstrated a simple technique to directly measure the blood flow speed in main arteries based on the diffused light model. The concept is demonstrated with a phantom that uses intralipid hydrogel to model the biological tissue and an embedded glass tube with flowing human blood to model the blood vessel. The correlation function of the measured photocurrent was used to find the electrical field correlation function via the Siegert relation. We have shown that the characteristic decorrelation rate (i.e., the inverse of the decoherent time) is linearly proportional to the blood speed and independent of the tube diameter. This striking property can be explained by an approximate analytic solution for the diffused light equation in the regime where the convective flow is the dominating factor for decorrelation. As a result, we have demonstrated a non-invasive method of measuring arterial blood speed without any prior knowledge or assumption about the geometric or mechanic properties of the blood vessels.

Keywords: blood flow; diffuse correlation spectroscopy; optical sensor.

MeSH terms

  • Arteries*
  • Blood Flow Velocity
  • Diagnostic Techniques, Cardiovascular
  • Hemodynamics*
  • Humans
  • Phantoms, Imaging