Intraocular lens power calculation with ray tracing based on AS-OCT and adjusted axial length after myopic excimer laser surgery

J Cataract Refract Surg. 2022 Aug 1;48(8):947-953. doi: 10.1097/j.jcrs.0000000000000902.

Abstract

Purpose: To report the results of intraocular lens (IOL) power calculation by ray tracing in eyes with previous myopic excimer laser surgery.

Setting: G.B. Bietti Foundation I.R.C.C.S., Rome, Italy.

Design: Retrospective interventional case series.

Methods: A series of consecutive patients undergoing phacoemulsification and IOL implantation after myopic excimer laser was investigated. IOL power was calculated using ray-tracing software available on the anterior segment optical coherence tomographer MS-39. Axial length (AL) was measured by optical biometry, and 4 values were investigated: (1) that from the printout, (2) the modified Wang/Koch formula, and (3) the polynomial equation for the Holladay 1 and (4) for the Holladay 2 formulas. The mean prediction error (PE), median absolute error (MedAE), and percentage of eyes with a PE within ±0.50 diopters (D) were reported.

Results: The study enrolled 39 eyes. Entering the original AL into ray tracing led to a mean hyperopic PE (+0.56 ±0.54 D), whereas with the Wang/Koch formula, a mean myopic PE (-0.41 ±0.53 D) was obtained. The Holladay 1 and 2 polynomial equations lead to the lowest PEs (-0.10 ±0.49 D and +0.08 ±0.49 D, respectively), lowest MedAE (0.37 D and 0.25 D), and highest percentages of eyes with a PE within ±0.50 D (71.79% and 76.92%). Calculations based on the Holladay 2 polynomial equation showed a statistically significant difference compared with other methods used (including Barrett-True K formula), with the only exception of the Holladay 1 polynomial equation.

Conclusions: IOL power was accurately calculated by ray tracing with adjusted AL according to the Holladay 2 polynomial equation.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biometry / methods
  • Humans
  • Lasers, Excimer / therapeutic use
  • Lens Implantation, Intraocular
  • Lenses, Intraocular*
  • Myopia* / surgery
  • Optics and Photonics
  • Phacoemulsification*
  • Refraction, Ocular
  • Retrospective Studies
  • Tomography, Optical Coherence