Pressure Ramp Testing for Optimization of End-Expiratory Pressure Settings in Adaptive Servo-Ventilation Therapy

Circ Rep. 2021 Nov 18;4(1):17-24. doi: 10.1253/circrep.CR-21-0132. eCollection 2022 Jan 7.

Abstract

Background: Clinical outcomes of adaptive servo-ventilation (ASV) therapy have not been rigorously assessed. Optimal device settings ascertained by a pressure ramp test may increase the utility of ASV therapy. Methods and Results: Patients with congestive heart failure (CHF) who underwent ASV therapy were prospectively included in the study. Patients in the ramp test group underwent a pressure ramp test, during which the end-expiratory pressure was optimized to maximize cardiac output (assessed using the AESCLONE mini). The control group consisted of age-matched patients who received ASV therapy with a default pressure 5 cmH2O. The primary endpoint was a composite of all-cause death and heart failure recurrence, and was compared between the 2 groups. Of a total of 37 patients, 11 each were included in the ramp test and control groups. Median patient age was 73 years (interquartile range 59-75 years) and 16 were men. There were no significant differences in baseline characteristics between the 2 groups. In the ramp test group, end-expiratory pressure was optimized between 2 and 5 cmH2O in each patient. The 2-year incidence of the primary endpoint tended to be lower in the ramp test than control group (0% vs. 59%; P=0.080). Conclusions: Pressure ramp testing may be a promising strategy to optimize device pressure settings in patients with CHF undergoing ASV therapy. Larger-scale trials are needed to validate our findings.

Keywords: Heart failure; Hemodynamics; Non-invasive positive pressure support.