Tracking Peripheral Artery Motion and Vascular Resistance With a Multimodal Wearable Sensor Under Pressure Perturbations

J Biomech Eng. 2022 Aug 1;144(8):081003. doi: 10.1115/1.4053399.

Abstract

The status of peripheral arteries is known to be a key physiological indicator of the body's response to both acute and chronic medical conditions. In this paper, peripheral artery deformation is tracked by wearable photoplethysmograph (PPG) and piezo-electric (polyvinylidene difluoride, PVDF) sensors, under pressure-varying cuff. A simple mechanical model for the local artery and intervening tissue captures broad features present in the PPG and PVDF signals on multiple swine subjects, with respect to varying cuff pressure. These behaviors provide insight into the robustness of cardiovascular property identification by noninvasive wearable sensing. This is found to help refine noninvasive blood pressure measurements and estimation of systemic vascular resistance (SVR) using selected features of sensor amplitude versus applied pressure.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Animals
  • Arteries
  • Hemodynamics
  • Humans
  • Photoplethysmography* / methods
  • Swine
  • Vascular Resistance
  • Wearable Electronic Devices*