Design and implementation of a highly integrated dual hemisphere capsule robot

Biomed Microdevices. 2022 Jan 19;24(1):10. doi: 10.1007/s10544-022-00611-5.

Abstract

To achieve cancer screening in any appointed position in 3D regions of the gastrointestinal (GI) tract such as esophagus, stomach and colon, a highly integrated dual hemisphere capsule robot (DHCR) with a novel three-layer nested structure is proposed. Based on tracking effect, in which the robotic axis is likely to be approximately coincident with the orientation of the space universal rotating magnetic field (SURMF) using the gyroscope dynamic balance, the dual hemisphere structure realizes the observation at a fixed-point in the passive mode and the rolling locomotion in the active mode by the dynamic posture control of the SURMF manipulation. The image acquisition module, wireless transmission module and driving actuator are tuned in a spherical structure, making the DHCR more compact and less invasive. To verify the maneuverability of the innovative DHCR both for observation at a fixed-point and navigation in curved intestine by aid of image, experiments are conducted in the simulated GI tract environment. The results show that the DHCR achieves effective conversion between posture adjustment and rolling locomotion, which lays a foundation for all-over inspection and medical operation inside 3D regions of the GI tract of human body.

Keywords: Dual hemisphere capsule robot (DHCR); Navigation in curved intestine; Observation at fixed-point; Space universal rotating magnetic field (SURMF); Tracking effect.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Capsule Endoscopy*
  • Equipment Design
  • Gastrointestinal Tract
  • Humans
  • Intestines
  • Magnetic Fields
  • Robotics*